Ph And Conductivity Connections; Ph/Orp/Pion Sensor Module Connections - ABB AWT420 Commissioning Instructions

Universal 4-wire, dual-input transmitter
Hide thumbs Also See for AWT420:
Table of Contents

Advertisement

AW T4 2 0 | U N I V E R S A L 4 - W I R E , D U A L- I N P U T T R A N S M I T T E R | C I/A W T4 2 0 - E N R E V. B

pH and conductivity connections

pH/ORP/pIon sensor module connections

NOTICE
ORP (Redox) and Antimony pH sensors do not feature
temperature compensation therefore do not have
temperature sensors or related wiring.
Standard sensors without diagnostic functions
NOTICE
Ensure sensor diagnostics are Off when using standard
sensors without diagnostic functions.
Sensor type
RTD wiring
2867
2-lead
TB5
2-lead
2-lead
AP1xx
3-lead
2-lead*
AP3xx
3-lead
APS1xx
2-lead*
APS5xx
3-lead
APS7xx
* Cut and remove grey wire
Standard sensors with diagnostic functions
NOTICE
Ensure sensor diagnostics are On when using standard
sensors with diagnostic functions.
Sensor type
RTD wiring
TBX5
2-lead
2-lead*
AP2xx
3-lead
* Cut and remove grey wire
SENSE
GUARD
1
2
Clear
Blue
Clear
Clear
Blue
Blue
Blue
Blue
SENSE
GUARD
1
2
Blue
Yellow
Black
Clear
Red
Blue
Clear
Red
Blue
REF
S .GND
RTD 1
3
4
Black
Black
Black
Black
Black
Black
Black
Black
REF
S .GND
3
4
Green
Green/Yellow
Green/Yellow
RTD 2
SHIELD
5
6
Red
White
Red
White
Red
White
Red
White
Red
Red
White
Red
White
Red
White
Yellow
Red
White
Yellow
RTD 1
RTD 2
SHIELD
5
6
Red
White
Dark green
Red
White
Red
White
13
RTD 3
7
8
Red
Grey
Grey
RTD 3
7
8
Grey