Array Script:
cs "REF BANDWIDTH INDEX", 8
cd "REF BANDWIDTH INDEX", 0, 2.5e01
cd "REF BANDWIDTH INDEX", 1, 5.5e01
cd "REF BANDWIDTH INDEX", 2, 1.25e02
cd "REF BANDWIDTH INDEX", 3, 3.0e02
cd "REF BANDWIDTH INDEX", 4, 6.5e02
cd "REF BANDWIDTH INDEX", 5, 1.55e03
cd "REF BANDWIDTH INDEX", 6, 3.65e03
cd "REF BANDWIDTH INDEX", 7, 1.05e04
cc "REF BANDWIDTH INDEX"
cs "REF BANDWIDTH LATCH VALUES", 8
cd "REF BANDWIDTH LATCH VALUES", 0, 0
cd "REF BANDWIDTH LATCH VALUES", 1, 1
cd "REF BANDWIDTH LATCH VALUES", 2, 2
cd "REF BANDWIDTH LATCH VALUES", 3, 3
cd "REF BANDWIDTH LATCH VALUES", 4, 4
cd "REF BANDWIDTH LATCH VALUES", 5, 5
cd "REF BANDWIDTH LATCH VALUES", 6, 6
cd "REF BANDWIDTH LATCH VALUES", 7, 7
cc "REF BANDWIDTH LATCH VALUES"
ky 58881
5. Press
Execute
.
Update the Arrays if Serial Number Prefix is ≥ US/MY 5150
1. Connect the instrument to a LAN and write down the IP address.
Press
>
Utility
Instrument Info/Help Mode
2. Load the IP address into the instrument.
Press
>
Utility
GPIB/RS-232/LAN
3. Using the web browser on the PC, enter the following URL, replacing <ip-address> with the actual IP address you
entered into the instrument in the previous step.
Type: http://<ip-address>/update.
4. Copy the script below and paste it into the text area on the System Update page shown in Figure 1.
Figure 2
System Update Web Page
>
.
Diagnostic
>
>
LAN Setup
IP Address
. Enter the IP address.