Download Print this page

Keysight Technologies E8257N Installation Note page 16

Psg signal generators add option unx ultra-low phase noise performance
Hide thumbs Also See for E8257N:

Advertisement

For Instruments with Motherboard Versions Less than A0101, replace A31L400 (E8257D) or
A31L44 (E8267D) (For Instruments with Serial Prefixes <US4805/MY4805)
Refer to Figure 6.
1. Position the signal generator with the A31 Motherboard facing you.
2. Remove A31L400 (E8257D) or A31L44 (E8267D).
3. Install the new inductor, provided in this kit.
NOTE
There is a spare replacement inductor provided in this kit.
Figure 6
16
A31L400 and A31L44
E8257D- L400
E8267D- L44
Front Panel