Environmental conditions
Optimum ambient temperature: Between 15 °C and 32 °C
Storage temperature: Between -15 °C and +60 °C (the well, pump and
the hoses must not contain any water)
Relative humidity: Between 30 % and 75 %
Air pressure: Between 794 hPa and 1060 hPa
Water and cooling conditions
Permissible water pressure (water inlet): 0.8 MPa (8 bar)
Maximum coolant temperature: 16°C
The device is not to be used outdoors.
The device is not suitable for operation in mobile facilities (ships, buses, trains
or oil platform). The unit has to be adapted in order to guarantee trouble-free
operation when used in mobile facilities.
Adaptation of the unit is required for operating altitudes greater than 2000 m
above sea level in order to guarantee trouble-free operation.
Please contact the customer service department without fail before putting the
device into operation - see page 2.
1.5.5

Commissioning

Initial commissioning will be undertaken by the manufacturer's own service
personnel or by an authorized service company. The device will be handed
over ready for use.
If the device has not been used for a longer period of time then it must first be
inspected by our customer service staff or by another authorized company
before being put back into operation again.
The device is only to be used if it is working correctly!
None of the safety devices (e.g. switch) or safety warnings are to be removed
or deactivated as this would impair the safe operation of the device!
The device must never be used if it is empty.
A water cooling system must be connected to the device if it is to be used for
more than 1 hour per day at maximum pressure. The pressure hoses must
not be removed. Please contact the customer service department without fail
before putting the device into operation - see page 2.
The device may only be operated with the pressure hose included in the
scope of delivery.
A water shut-off valve must be used.
High frequency mobile communication systems can interfere with the way the
device functions.
Heat therapy devices (microwave therapy devices, etc.) installed in the vicinity
may lead to electromagnetic interference. Contact our customer service de-
partment if this happens – see page 2.
The device must be connected to the mains via an all pole master switch.
One power supply circuit (power supply lead) may only power one device.