Connecting The Cables; Correct Positioning Of The Cables; Operating Instructions; Operation Of The Height Adjustment And Profiling Functions - Park House Healthcare BUTTERFLY COT User Manual

Table of Contents

Advertisement

4.6 Connecting the cables

The Control box has colour coding to clearly indicate the correct cable connections.
IMPORTANT – Please ensure that all cables are connected correctly, and test all of the functions fully
before using the cot
Only connect the components as shown.
Connecting differently may result in damage to the control unit.

4.7 Correct positioning of the cables

The cables must be connected in the correct manner to avoid trapping the cables between the metal parts of the cot
during use. This could cause damage to the cables and create a serious safety hazard.
IMPORTANT – Before using the cot, and after re-locating the lock-out box, check that all cables have been fitted
correctly and that they can not become trapped, damaged and dangerous during use.
5.

Operating instructions

IMPORTANT - Any person using this cot must read this instruction manual carefully and they should be fully
conversant with the correct operating procedures for its safe operation.

5.1 Operation of the height adjustment and profiling functions

The head section, knee break, cot height and Trendelenburg tilting are adjusted by pressing the appropriate button on
the handset. For ease of use, the buttons are labelled graphically according to their function.
WARNING - Do not exceed the maximum usage period – all functions maximum of 6 minutes per hour.
Colour Codes
1. Green: Lock-out device.
2. Red: Handset
3. Black: Backrest motor.
4. Yellow: Footrest motor.
5. Blue: Head raise-up motor
6. White: Foot raise-up motor
7. Black: Backup battery plug.
7