Calibration - ResMed S8 Compact Clinician Manual

System
Hide thumbs Also See for S8 Compact:
Table of Contents

Advertisement

Available languages

Available languages

C
ALIBRATION
The
S8 C
calibration for home use. If there has been a change in altitude within a given altitude
compensation range, or a change in circuit configuration, we recommend you check
the calibration and adjust as required.
Note: Some countries require all medical equipment to have a periodic service and
calibration. Please check the regulations in your country.
Q
C
UICK
To check the calibration of the flow generator:
1. In the Clinical Menu, set the CPAP pressure setting to 12 cm H
2. Measure the pressure at the mask using an accurate manometer. If the delivered
pressure is less than 11.5 cm H
recalibrate the flow generator.
A
DJUSTING THE
To calibrate the
of 0 to 22 cm H
(0.098" to 0.118"/2.5 mm to 3.0 mm internal diameter) and a standard Luer taper
fitting.
1. Assemble the
tubing and the mask system. Connect all accessories (humidifier, filters, etc) that
are normally used.
2. Turn on the
start running, press the Stop/Start key to start the airflow. Let the device warm
up for 15 minutes.
3. While the device is warming up, remove the plug from one of the access ports
on the mask, and attach the Luer taper end of the sensor tube. The other end
of the sensor tube should be fitted to the manometer. If you have a blind
B
UBBLE
4. When the motor has been running for 15 minutes, press the Stop/Start key to
stop the airflow.
5. Access the Clinical Menu from the
Down and Right keys simultaneously for at least 3 seconds.
6. Scroll to the Altitude Compensation screen (ALT). Ensure that the range is set to
the appropriate altitude setting. If necessary, use the Up and Down keys to set
the appropriate range.
7. Scroll to the CPAP menu.
8. Press the Left key to change the pressure setting. Use the Up and Down keys to
set the pressure to 12 cm H
pressure.
9. Scroll to the Calibrate screen. Press the Left key (Change) to change the
calibration setting. The airflow will start.
24
is factory calibrated before delivery and will not require periodic
OMPACT
HECK
C
ALIBRATION
S8 C
OMPACT
O in intervals of 0.1 cm H
2
S8 C
OMPACT
S8 C
OMPACT
C
, attach it to the mask.
USHION
TM
O or more than 12.5 cm H
2
you will need a reference manometer with a range
O, fitted with a sensor tube
2
as the patient would normally use it, complete with air
to allow the device to warm up. If the motor does not
S8 C
OMPACT
O. Press the Left key (Change) to apply this
2
O.
2
O, you may need to
2
screen by holding down the