Cautions - ResMed S8 Compact Clinician Manual

System
Hide thumbs Also See for S8 Compact:
Table of Contents

Advertisement

Available languages

Available languages

The
event of certain fault conditions, pressures up to 30 cm H
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
If oxygen has been left on, turn off the flow generator, then wait 30 minutes
before turning on the flow generator again.
Explanation: When the CPAP device is not in operation and the oxygen flow is
left on, oxygen delivered into the air delivery tubing may accumulate within the
CPAP machine enclosure and create a risk of fire. This applies to most types of
CPAP machines.
Note: At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on where the oxygen is introduced, the pressure settings, patient
breathing pattern, mask selection, and leak rate.
Do not use the
changes in performance.
Do not open the
Repairs and internal servicing should only be performed by an authorized service
agent.
Note: The plate on the back of the device is a non removable cover. Removal of this cover
voids warranty.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
C
AUTIONS
At low pressures, the flow through the exhalation ports of your mask may not
clear all exhaled gas from the tubing. Some rebreathing may occur.
The air flow for breathing produced by this device can be as much as 11°F (6°C)
higher than the temperature of the room. Caution should be exercised if the room
temperature is warmer than 90°F (32°C).
Note: The above are general warnings and cautions. Specific warnings, cautions, and
notes appear with the relevant instructions in the manual.
6
S8 C
can be set to deliver pressures up to 20 cm H
OMPACT
Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
Always ensure airflow is being generated by the device before the oxygen
supply is turned on.
Always turn the oxygen supply off before stopping the airflow from the device.
S8 C
OMPACT
S8 C
OMPACT
if there are obvious external defects or unexplained
case. There are no user serviceable parts inside.
O. In the unlikely
2
O are possible.
2