ResMed S8 Compact Clinician Manual page 36

System
Hide thumbs Also See for S8 Compact:
Table of Contents

Advertisement

Available languages

Available languages

S
S
YSTEM
PECIFICATIONS
S8 C
OMPACT
Performance
Operating pressure range: 4 to 20 cm H
Dimensions (H x W x D): 4.6" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight: 2.9 lb (1.3 kg)
Power Supply
Input range for
2.5A < 140VA (110W) (maximum power consumption). Instantaneous peak power
consumption <340 VA.
Actual power consumption will vary, depending on factors such as the use of
accessories, height above sea level, and ambient temperature. Power consumption
values for typical treatment conditions (using a breathing machine set to 0.5 L and
15 breaths/min with 6'6" (2 m) air tubing and an
following table.
Device
S8 C
OMPACT
Power outlet for HumidAire 3i
Outlet range: 100–240V, 50/60Hz; 110V, 400Hz; 2.0A
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6'6" (2 m)
Air Outlet: The 22 mm conical outlet complies with EN 1281-1
IEC 60601-1 Classifications: Class II (double insulation), Type CF
Noise Level: Radiated sound pressure is measured at <30dBA at 10 cm H
3'3.36" (1 m ) according to ISO17510-1.
Housing Construction: Flame retardant engineering thermoplastic
Environmental Conditions
Operating Temperature: +41°F to +104°F (+5°C to + 40°C)
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -4°F to +140°F (-20°C to 60°C)
Storage and Transport Humidity: 10%–95% non-condensing
Operating Altitude: Sea level to 8500' (2591 m)
S8 C
H
with
OMPACT
UMID
Pressure
(cm H
O)
2
5
10
15
20
O
2
A
3i
:100–240V, 50/60Hz; 110V, 400Hz;
IRE
U
M
LTRA
IRAGE
Power
(VA)
17
23
29
34
mask) are given in the
Power
(W)
8
12
16
20
O and
2
S
S
YSTEM
PECIFICATIONS
31

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents