You can set a timed reminder to remind your patient when to return the device for
service. The patient can press the Left (ok) key to remove the message from the LCD.
Table 4 shows the options and default settings for the types of reminders in the
Reminder Menu.
Table 4: Reminder type options and default settings
Type of reminder
Replace mask (MASK)
Call provider (CALL)
Insert Card (CARD)
Change filter (FILT)
Service due (SERV)
*
OFF means that the date set for the reminder is today's date or earlier.
This feature is not available.
A
C
LTITUDE
Changes in altitude will affect the pressure delivered by the
has a feature which compensates for changes in altitude.
The altitude compensation feature can be set to one of four altitude ranges. During
calibration, a change to the pressure gain within an altitude range will provide a more
accurate delivered pressure.
Figure 5 shows the effect of altitude (with compensation) on pressure when set to
10 cm H
ambient temperature of 68°F (20°C) and 50% relative humidity.
* Altitude setting change
Figure 5: Effect of altitude on pressure
22
D
UE
Default setting
*
OFF
OFF
OFF
OFF
OFF
OMPENSATION
O, and where the
2
Mask Pressure vs Altitude
13
12
11
10
*
9
8
7
0
2000
Options
7-day increments, starting from current date
7-day increments, starting from current date
7-day increments, starting from current date
7-day increments, starting from current date
7-day increments, starting from current date
S8 C
is calibrated at sea level 0 ft (0 m) at an
OMPACT
(S8 calibrated for 10 cm H
*
4000
6000
Altitude (ft)
S8 C
OMPACT
O at 0')
2
*
8000
. The device
10000