ResMed S8 Compact Clinician Manual page 24

System
Hide thumbs Also See for S8 Compact:
Table of Contents

Advertisement

Available languages

Available languages

Table 2: Parameters in the Patient and Clinical Menus
Parameter
Altitude
*
Compensation
*
Ramp
Max Ramp
Start CPAP
CPAP
Calibrate
*
Used Hrs
*
Usage
Erase
Language*
Time
Date
SN
SW*
Motor Hours
*
Available on both the Patient Menu and the Clinical Menu.
Refer to the S8 Compact Quick Set-up Guide for details of factory settings.
Description
Sets the altitude compensation range. See "Altitude
Compensation" on page 22.
Selects the Ramp time. (Available in standby mode.)
Sets the maximum ramp time the patient can select.
Sets the pressure at the start of the ramp time up to fixed
treatment pressure.
Sets the fixed treatment pressure.
Adjusts the measured pressure at the manometer to
match the set CPAP pressure. Use if there is a differential
of greater than 0.5 cm H
pressure and the measured pressure.
Changing the calibration will change the pressure gain
displayed on the LCD.
Entering the Calibrate screen will start airflow. Exiting
this screen will stop airflow.
See "Calibration" on page 24.
Displays the total number of treatment hours. (View
only.)
The number of used sessions greater than 1 minute in
length over the number of days elapsed since the first
used day.
Resets the used hours and used days.
The language in which menus and settings are displayed
on the LCD.
Sets the current time.
Displays the current date.
Displays the serial number of the device.
Displays the current software version.
Displays the total number of hours the S8 Compact
motor has been running, to a maximum of 18,250 hours.
O between the set CPAP
2
N
S8 C
AVIGATING THE
19
M
OMPACT
ENUS

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents