Installing/mounting

2.1

General safety instructions

Process pressure
Improper installation may result in loss of process pressure.
Chemical resistance against the medium
Materials of construction are chosen based on their chemical compatibility (or inertness) for
general purposes. For exposure to specific environments, check with chemical compatibility
charts before installing.
Temperature range
The range of the ambient and process temperature of the device must be observed (see
Ambient Temperature (Page 39)and for Ex-approvals here). (Page 12)
Special conditions of use
Model 7ML5738 Extension tube is constructed using Industry Electrical conduit installed by
qualified personnel per equipment manual and NFPA 70 National Electrical Code (FM
Approved Unit).
Model 7ML5738 Extension tube is constructed using Industry Electrical conduit installed by
qualified personnel per equipment manual and CEC C22.1 Canadian Electrical Code.
Mechanical load
The torque at the fastening spot must not exceed 180 Nm (Pipe extension)/ 100 Nm
(Customer supplied pipe extension)
Maximum length "L" in dependence on the deviation (in degrees) from vertical installation
=Load
SITRANS LVS300
Operating Instructions, 05/2019, A5E44908583-AB
CAUTION
CAUTION
CAUTION
CAUTION
CAUTION
2
7