Before Use

Please read through the instructions
carefully before using the unit and
carefully note the safety advice.
Always keep this manual accessible
to any users of the drill.

Area of application

This device should only be used by
people who have completed a foot
care qualification. Before use the
operator should familiarize with the
device by reading this manual or
being instructed by another.
The device must only be used in
closed spaces. It may be used in
hospitals, in a commercial environ-
ment or in homes. The device is not
intended for use in operating rooms.
The control unit can be set up in a
cabinet, or can be used as a porta-
ble unit (e.g. in a carry case). It is
suitable for the following scopes:
Foot care: Milling, trimming and
polishing
.
Attention:
must be in operation during foot
care.
• Manicure: Grinding, milling and
polishing works.
The devise is determined of
operating rotating instruments,
which will not penetrate the
skin or murcous membrane if
used appropriate, but could
evoke damages accidently.
Caution: In the area of pedicure
you have to work with spray
techniques.
The
spray
9

Operation

The working area must conform to
the conditions described. Set up the
unit in such a way, that the air slits
are not blocked and the extracted
air can escape well. To guarantee
unhindered air flow there must a
gap of at least 12 cm to the right of
the device. Position the device so
that the operating buttons are easy
to reach and the outlet of the hose
is a maximum distance of 60 cm
from the patient's foot. The unit
should be 30 – 90 cm from the floor.
Image: Example of positioning of foot care unit
in treatment room.
Image: Bird's eye view of work area