Safety Instructions
For safety reasons you must always wear disposable gloves
and protective glasses when operating this unit. In addition, a
face mask conforming to EN 14683 type IIR or EN 149 FFP 2
must also be worn.
If water leaks from the unit, then a disturbance is present. The
unit has to be separated from the socket and cannot be used
any longer.
Ensure the patient's foot and the hand piece (e.g. material) are
held firmly in place during operation.To avoid injury, work on the
foot of the patient with extreme caution, and light pressure.
Only use distilled water plus a sterilizing additive, e.g. hadewe
AquaHy (art. no. 8555). If you use other brands, pay attention to
their manuals if you can use them for this purpose. Never use
tap water or alcoholic fluid, or other burnable liquids. In case,
you are working with an alcoholic fluid, or other burnable liquids,
there is always the danger of fire or explosion hazards. Never
refill the tank directly over the unit to avoid in case you
accidently spill water, this will flow into the unit.
Change the spray liquid everyday. Always add disinfection
additive to the distilled water. Rinse the tank daily and fill it with
fresh water everyday.
To ensure that the bur/tool sits firmly in the chuck, do not work
with tools which:
have an oily shaft
are worn
are bent.