Load cell connecting definition (5pin aviation waterproof connector), see the area in the
circle:
 
 
1pin:(E+)red cable
2pin:(E-)black cable
3pin:(SH)shield cable
4pin:(S+)green cable
5pin:(S-)white cable
 
5.0

CALIBRATION

The PTS plus is factory calibrated, if further information is required for re-calibrating the
scale the procedure is as follows:
Turn on the power and wait for the scale to configure.
Before configuration is completed, hold down the [Tare] ↵ key to bring up the
user password information screen.
To enter the password, use the directional buttons ⭡ or ⭣ to scroll through
numbers 1-9. Use the ⭢ directional key to move to the next digit.
Entering the correct password [1000] and pressing [Tare] for confirmation
will bring you to the user setup menu;
Within the menu, use the directional buttons ⭡ or ⭣ to scroll through the
settings until 'U-cal' appears on the display. Press the [Tare] ↵ key to select.