Summary; Appendix; Technical Data - Greiner Bio-One CheckScanner 2.0 Instructions For Use Manual

Table of Contents

Advertisement

17. SUMMARY

Your
Power
for Health
The CheckScanner™ 2.0 is a PC-controlled biochip fl uorescence scanner for the daily general
laboratory use in combination with the CheckReport™Software.
It is important that the operator reads this manual thoroughly before using the instrument. The
scanner must be operated by professionals only. The service technician (GBO) must install and
initialize the device (installation qualifi cation). After that, a test run must be carried out which ensures
the correct operation. Hazardous samples must not be measured. Since no infectious samples are
measured, no special disinfection is necessary.
The CheckScanner™ 2.0 is designed to scan with high performance ready-to-use microarrays on
HTA™Slides manufactured by Greiner Bio-One. It is important that the operator reads this manual
thoroughly before using the instrument. The scanner must be operated by professionals only. The
manufacturer does not assume any liability for any use of the instrument not described in this manual,
or for any customer modifi cations made to the instrument. Subsequent scans of the microarray are
performed at two different wavelengths. The output of the instrument are TIFF images which are
evaluated using the CheckReport™Software.
The independent scanning of up to four HTA™Slides is automatically carried out in one scan job
according to specifi c scan protocols.

18. APPENDIX

18.1

Technical Data

The CheckScanner™ 2.0 is a PC-controlled biochip fl uorescence scanner for the daily general
laboratory use in combination with the CheckReport™Software.
Item
Sample Size/Capacity
Laser Excitation Wavelengths
Emission Filter Wavelengths
Laser Safety Class according
EN 60825-1:2014
Embedded Laser Class according
EN 60825-1:2014
Focus
Simultaneous Wavelengths per Scan
Sample/Substrate Dimension
Sample Loading
Scanning Field
Pixel Resolution
Scan Speed (per wavelength)
Dynamic Range
Greiner Bio-One GmbH
Maybachstr. 2
72636 Frickenhausen
Germany
Made in Germany
CheckScanner™ 2.0 - Instructions for Use
Revision: BQ-686-01 /March 2017
Specifi cation
Up to 4 HTA
532 nm / max. 10 mW and 639 nm / max. 20 mW (solid-state lasers)
570 nm and 670 nm (standard)
Easy exchange with two positions fi lter drawer or 6 positions fi lter wheel
1
3B
Motorized focus adjustment
One
ISO slide standard: max. width 25.4 mm; min. width 24.9 mm (up to 4
simultaneously)
Force-free loading of HTA
23 x 76mm
10µm
a) 20 x 20mm area: approx. 3.5 min
b) 22 x 73mm area: approx. 13 min
16-bit dynamic range, linear over 4 decades
862 070
Slides per run
TM
Slides
TM
19

Advertisement

Table of Contents
loading

Related Products for Greiner Bio-One CheckScanner 2.0

Table of Contents