S
Evaluating Results - Concepts
Instantaneous Measurements
Instantaneous Measurements - Concepts
Agilent J-BERT N4903B High-Performance Serial BERT
User Guide
7
Evaluating Results
The Serial BERT offers several different kinds of tests that can be run:
•
Instantaneous Measurements
This type of measurements is used to monitor the instantaneous "BER" / "8B/
10B Comparison" / "Bit Comparison without PCIe3 SKPOS" / "Bit Comparison
without USB3.1 SKPOS" during measurement setup, experiments, and
adjustments.
•
Accumulated Measurements
This type of measurements is used to measure and log error performance over
time. Also, you can capture the location of bit errors for further analysis.
•
Eye Measurements
This type of measurements is used to measure the eye height and width of the
incoming data signal at specific alignment BER thresholds. This information is
displayed on a representation of an eye diagram.
With instantaneous measurements, you can directly view the current "BER" /
"8B/10B Comparison" / "Bit Comparison without PCIe3 SKPOS" / "Bit
Comparison without USB3.1 SKPOS" results on the screen.
The "8B/10B Comparison" includes the following measurements:
•
Symbol Error Ratio (SER)
s