prepared reagent solutions can be stored for 3 months, after which time the canisters
and remaining contents must be discarded.
1. Pour additional component A into the reagent A canister using the funnel.
2. Pour additional component D into the reagent D canister using the funnel.
3. Close reagent canisters A and D using new caps and shake each one thoroughly for
4. Stick one label displaying the contents on each of the reagent canisters A and D.
5. Press a menu button for three seconds and select the
6. Open the refrigerator and remove the reagent hoses from the canisters.
7. Remove the old reagent canisters from the refrigerator, unscrew the caps and replace
8. Place the new canisters in the refrigerator, connect the hoses, close the refrigerator,
Replacement of standard solution
The standard solution is ready to use. Replace the standard solution every 6–12 months,
depending on use.
1. Remove the old bottle of standard solution, unscrew cap and replace with cap of new
2. Insert new bottle, screw cap onto intake hose.
3. The exact concentration of the standard solution is printed on the bottle.
Enter the standard concentration in the
Replacement parts & reagents
Description
Reagents A, B, C and D and additional components
Annual wear parts set
Reagent A, 5 L canister
Additional component A, 250 mL wide-mouth bottle
Reagent B, 5 L canister
Reagent C, 5 L canister
Reagent D, 5 L canister
Additional component D, 100 mL wide-mouth bottle
Standard solution, 500 mL square bottle
14
one minute.
them with the caps from the new canisters.
then select
PREPUMPING
bottle.
and then
.
CALIBRATION
+SETTINGS> STANDARD MENU.
menu.
+SERVICE
Order no.
LCW893
LZV809
BCF689
BCZ822
BCF1161
BCF691
BCF692
BCZ824
LCW824