User interface

User interface
The menu items
menu in the user interface. All other menus and user operations have remained
unchanged in the move from Phosphax sigma to Phosphax sigma HR and will not be
explained in this supplement. Further information can be found in the PHOSPHAX sigma
user manual DOC023.53.03113.
Menu
All instrument parameters for adjustment to the appropriate conditions are defined in the
+SETTINGS
out and the current output is updated. If no buttons are pressed for over ten minutes, the
instrument returns automatically to the measurement operation display.
Designation
Function
CONTRAST
Readability of display
LANGUAGE
Selection of menu language
STANDARD
Concentration of standard solution in use
+SPECIAL P
Submenu for total phosphorus settings
+SPECIAL OP
Submenu for orthophosphate settings
Number of measurement cycles per
P/OP
parameter
RELAY-MIN
Lower limit contact
RELAY-MAX
Upper limit contact
SIGMATAX
Sample feed via SIGMATAX
MEAS. INTERVAL
Time between two measurements
Wait time before a measurement,
MEAS. DELAY
to rinse pipes etc.
DATE
Date setting
TIME
Time setting
CLEAN
Automatic cleaning
CALIBRATION
Automatic calibration
VERSION
Program version
PASSWORD
Password protection on activation
SW-LOAD
EPROM type
SERVICE
Place and contact
STATION
LIGHTING
Display illumination
12
MEAS. INTERVAL
+ SETTINGS
menu. While the instrument is in this menu, further measurements are carried
and
have been added to the +
MEAS. DELAY
Available settings
-90 to +90
DE, GB, FR, NL
1.95–2.05 in 0.1 increments
FACTOR, CUR. RANGE, CURRENT, FAULT
FACTOR, CUR. RANGE, CURRENT, FAULT
0–10 each
0.00 to CUR. OUT P setting
0.00 to CUR. OUT P setting
Yes, no
15 to 240 minutes in 5 minute steps, remote, off
0 ... 600 minutes
MM:DD:YY
HH:MM, am, pm
Off, 6 h, 12 h, 1–7 days, start time always on the hour
->LAST Last clean displayed
Off, 1–7 days, start time always on the hour
->LAST Last calibration displayed
4 characters using digits 1–4
Yes, no
On, 20 minutes (after pressing button), time 5°°-17°°
SETTINGS
English

Advertisement