Time Base Accuracy Test; Specification; Test Connection Diagram; Test Procedures - Rigol DS6104 Performance Verification Manual

Digital oscilloscope
Hide thumbs Also See for DS6104:
Table of Contents

Advertisement

RIGOL

Time Base Accuracy Test

Specification

Time Base Accuracy
Specification
[1]
Note
: Typical.
[2]
Note
: Clock drift is lower than or equal to ±2 ppm/year.
[3]
Note
: For the number of years that the instrument has been used, please calculate according to the date in the
verification certificate provided when the instrument leaves factory.

Test Connection Diagram

Fluke 9500B

Test Procedures

1.
Connect the active signal terminal of Fluke 9500B to CH1 of the oscilloscope, as shown in the
figure above.
2.
Enable Fluke 9500B and set its impedance to 50 Ω
3.
Output a sine waveform with 10 MHz frequency and 1 Vpp amplitude via Fluke 9500B.
4.
Configure the oscilloscope:
1) Press CH1 in the vertical control area (VERTICAL) at the front panel to enable CH1.
2) Press CH1  Probe  Ratio to set the probe attenuation ratio to "1X".
3) Press CH1  Input to set the input impedance to 50 Ω.
4) Rotate VERTICAL
5) Rotate VERTICAL
6) Rotate HORIZONTAL
Tip: To quickly set the horizontal position to 1 ms, you can first rotate HORIZONTAL
2-18
[1]
≤ ±(4 ppm + Clock Drift
been used
)
[3]
Figure 2-5 Time Base Accuracy Test Connection Diagram
SCALE to set the vertical scale to 200 mV/div.
POSITION to set the vertical position to 0.
POSITION to set the horizontal position to 1 ms.
Chapter 2 Performance Verification Test
× Number of years that the instrument has
[2]
DS6000 Performance Verification Guide