To clean the EzSensor, use either of the following solutions listed below and observe the
precautions noted. (Do not use any type of solvent, such as alcohol or benzene.)
●
Mild soap and water
●
Isopropyl alcohol (70%)
●
Most alcohol and ammonia based cleaners
●
Mild, non-abrasive cleaners
Do not soak or immerse the system and be sure to dry it completely after cleaning.
Clean the surface of the system by moistening it with a soft cotton swab dipped in either of the
cleaning solutions listed above. Gently wipe the surface from end-to-end in straight lines,
without applying any pressure. Make sure that liquid does not penetrate the system through the
USB cable or the sensor cable connectors.
After cleaning the surface of the EzSensor, use a clean lint-free cloth to dry the system, as
required, until the surface is clean.
※Clean the silicone cover using the same method.
※ Do not use the following cleaning materials.
●
Hard brushes or scrapers of any kind
●
Strong acids or alkaloids
4.4
Precautions
●
Do not soak the sensor in the water or alcohol.
●
Authorized service personnel can repair about calibration.
●
Service personnel cannot handle the problem which is not mentioned in this manual.
So, in this case please request the repair to manufacturer through VATECH dealer.
●
This equipment and accessories are to be disposed safely at the end of the product life
cycle and national regulation must be observed.
4.5
Product complaint
Any health care professional (e.g., customer or user of this system of products) who has any
complaints should notify your distributor first. They will handle them.