Wiring procedure
Perform the following procedure to wire circuitry to the
screw terminal connector card:
WARNING
Make sure all power is off and that any
stored energy in external circuitry is dis-
charged.
1. If mated together, separate the connector card from the
relay card by removing the mounting screw and pulling
the two cards away from each other. Remember to han-
dle the cards only by the edges and shields to avoid con-
tamination.
2. Using an insulated screwdriver, connect the circuitry to
the appropriate terminals. Figure 3-6 shows how the out-
put of Bank A would be connected to a DMM.
3. Referring to Figure 3-7, remove the top half of the cable
clamp as follows:
A. Loosen the cable clamp screw enough to disengage
it from the bottom half of the cable clamp.
B. Using your thumb and forefinger, press the retaining
clips inward and, with your other hand, remove the
top half of the clamp.
4. Route wires under wire guide/connector shim.
5. Route the wires through the bottom half of the cable
clamp.
6. Replace the top half of the clamp. It simply snaps onto
the bottom half of the clamp. Tighten the cable clamp
screw. The clamp serves as a strain relief for terminal
block wires.
7. Mate the connector card to the relay card. The Model
7015 is now ready to be installed in the Model 7001 or
Model 7002 mainframe. See paragraph 3.5 for details.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Card Connections & Installation
#16 - 22 AWG Wires
Bank A
H
Output
L
H
CH 1
L
H
CH 2
L
H
CH 3
L
Figure 3-6
Typical screw terminal connections
Clips
Figure 3-7
Cable clamp for screw terminal connector card
HI
DMM
LO
Screw