Keithley 7016A Instruction Manual
Keithley 7016A Instruction Manual

Keithley 7016A Instruction Manual

50o 2ghz multiplexer card

Advertisement

Quick Links

Instruction Manual
Model 7016A
50 Ω 2GHz Multiplexer Card
Contains Operating and Servicing Information
7016A-901-01 Rev. A / 5-99

Advertisement

Table of Contents
loading

Summary of Contents for Keithley 7016A

  • Page 1 Instruction Manual Model 7016A 50 Ω 2GHz Multiplexer Card Contains Operating and Servicing Information 7016A-901-01 Rev. A / 5-99...
  • Page 2 WARRANTY Keithley Instruments, Inc. warrants this product to be free from defects in material and workmanship for a period of 1 year from date of shipment. Keithley Instruments, Inc. warrants the following items for 90 days from the date of shipment: probes, cables, rechargeable batter- ies, diskettes, and documentation.
  • Page 3 Model 7016A 50 Ω 2GHz Multiplexer Card Instruction Manual ©1999, Keithley Instruments, Inc. All rights reserved. Cleveland, Ohio, U.S.A. First Printing, May 1999 Document Number: 7016A-901-01 Rev. A...
  • Page 4 Revision of the manual. Each new Revision includes a revised copy of this print history page. Revision A (Document Number (7016A-901-01) ..................May 1999 All Keithley product names are trademarks or registered trademarks of Keithley Instruments, Inc. Other brand and product names are trademarks or registered trademarks of their respective holders...
  • Page 5 (IEC) Standard IEC 664, digital multimeter measuring circuits The types of product users are: (e.g., Keithley Models 175A, 199, 2000, 2001, 2002, and 2010) are Installation Category II. All other instruments’ signal terminals are Responsible body is the individual or group responsible for the use Installation Category I and must not be connected to mains.
  • Page 6 NOT as safety earth ground connections. leads, and input jacks, must be purchased from Keithley Instru- ments. Standard fuses, with applicable national safety approvals, If you are using a test fixture, keep the lid closed while power is ap- may be used if the rating and type are the same.
  • Page 7 7016A 50Ω 2GHz Multiplexer MULTIPLEXERS PER CARD: 2 (with isolated ground). CHARACTERISTIC IMPEDANCE: 50Ω nominal. 7016A Insertion Loss, Typical CHANNELS PER MULTIPLEXER: 4. LOG MAGNITUDE REF= 0.000dB 0.500 dB/DIV CONTACT CONFIGURATION: 1 pole Form A common shield. 0.0dB RELAY DRIVE CURRENT: 120mA.
  • Page 8: Table Of Contents

    Table of Contents General Information Introduction ................................1-1 Features ................................1-1 Warranty information ............................1-1 Manual addenda ..............................1-1 Safety symbols and terms ........................... 1-1 Specifications ..............................1-1 Unpacking and inspection ........................... 1-2 1.7.1 Inspection for damage ..........................1-2 1.7.2 Handling precautions ..........................
  • Page 9 Switching considerations ............................. 3-2 3.5.1 Card characteristics............................ 3-2 3.5.2 Cable characteristics ..........................3-3 Applications................................. 3-3 3.6.1 Multiplexer expansion ..........................3-3 3.6.2 Filter testing ............................... 3-4 3.6.3 Impedance testing ............................3-6 Service Information Introduction ................................. 4-1 Handling and cleaning precautions........................4-1 4.2.1 Handling precautions ..........................
  • Page 10 List of Illustrations Connections and Installation Figure 2-1 Model 7016A simplified schematic ......................2-2 Figure 2-2 Model 7016A configuration......................... 2-2 Figure 2-3 Typical connection scheme .......................... 2-3 50 Ω terminator installation .......................... 2-4 Figure 2-4 Figure 2-5 Card installation in Model 7001........................2-5 Figure 2-6 Card installation in Model 7002........................
  • Page 11 List of Tables Service Information Table 4-1 Recommended verification equipment ......................4-2 Table 4-2 Insertion loss values ............................. 4-2 Table 4-3 Recommended troubleshooting equipment....................4-8 Table 4-4 Troubleshooting procedure .......................... 4-8 Replaceable Parts Table 5-1 Model 7016A parts list..........................5-2...
  • Page 12: General Information

    This section contains general information about the Model Any improvements or changes concerning the card or manu- 701A6 50 Ω 2GHz Multiplexer Card. The Model 7016A is al will be explained in an addendum included with the card. equipped with two four-channel multiplexers and is designed for 50 Ω...
  • Page 13: Unpacking And Inspection

    Unpacking and inspection Repacking for shipment 1.7.1 Inspection for damage Should it become necessary to return the Model 7016A for repair, carefully pack the unit in its original packing carton The Model 7016A is packaged in a re-sealable, anti-static or the equivalent, and include the following information: bag to protect it from damage due to static discharge and •...
  • Page 14: Connections And Installation

    2.3.2 Card configuration This section includes information on making connections to the Model 7016A and installing the card in the Model 7001/ Figure 2-2 shows the general configuration of the Model 7002 Switch System. 7016A. Connectors include:...
  • Page 15 Ch. 4 Connector Ch. 5 Connector Ch. 6 Out B Multiplexer B Ch. 7 Ch. 8 Figure 2-1 Model 7016A simplified schematic Multiplexer B Multiplexer B Multiplexer A Multiplexer A SMB Terminator SMA Input/Output SMB Terminator SMA Input/Output Connectors Connectors...
  • Page 16: Input/Output Connecting Cables

    Figure 2-3 shows a typical connecting scheme for the Model should be made using 50 Ω coaxial cable equipped with 7016A. In this arrangement, sources are connected to the inputs SMA connectors. The recommended cable type is RG223/U. while the measuring instrument is connected to the outputs.
  • Page 17: Input Termination

    User-supplied 50 Ω female SMB terminators may be in- This paragraph explains how to install and remove the Model stalled for applications requiring proper 50 Ω termination of 7016A card assembly from the Model 7001/7002 main- off channels. Figure 2-4 shows typical installation of a 50 Ω frame.
  • Page 18: Card Installation

    Connections and Installation 2.4.1 Card installation 2.4.2 Card removal Perform the following steps, and refer to Figure 2-5 or Figure Follow the steps below to remove the multiplexer card from 2-6 to install the card assembly in the Model 7001/7002 the mainframe: mainframe.
  • Page 19 Connections and Installation Ejector Arms (locked position) WARNING: INTERCONNECTION, INSTALLATION AND REMOVAL OF CARDS BY QUALIFIED SERVICE PERSONNEL ONLY. CARD CARD CARD CARD WARNING: NO INTERNAL OPERATOR SERVICEABLE PARTS, SERVICE BY QUALIFIED PERSONNEL ONLY. Figure 2-6 Card installation in Model 7002...
  • Page 20: Operation

    Operation Introduction Model 7016A installed in slot 2, key in the following channel list, and press CLOSE: This section contains basic information on using the Model SELECT CHANNELS 2!4 7016A including signal limitations and switching consider- ations. For detailed mainframe operating information, refer To open a closed channel, simply press OPEN or OPEN to the Model 7001 or Model 7002 Instruction Manual.

Table of Contents