Page 1
Pocket Diagnosic Instr uments ® Use, Care and Maintenance ®...
Page 2
C UTION: Federal law restricts this device to sale by or on the order of a physician or licensed health- care practitioner. W RNING: The carrying cases provided with these devices are intended for long term storage or transport between facilities or when shipping devices to and from ADC’ s facility for repair or servicing. Carry cases...
Page 3
(including internal liners) can NOT be cleaned or disinfected and should be safely discarded if contaminated. Do not carry the case into a contaminated environment. W RNING: When replacing lamp, allow lamp to cool for five minutes before handling. W RNING: No form of sterilization has been validated for this device or its corresponding components and sterilization may compromise the safety and effectiveness of this device.
Page 4
NOTE: Allow lamp to cool for five minutes before handling if device has been recently used. (Otoscope lamp p/n 5111N-4, Ophthalmoscope lamp p/n 5112N-4) Battery Replacement This unit requires 2 “AA” batteries. For best performance, we suggest alkaline batteries.
Page 5
Otoscope and ccessories Troubleshooting Guide Trouble rea Possible Cause Corrective ction Lamp burned Replace the lamp with ADC part #5111N-4 No light output Wrong lamp Replace the lamp with installed ADC part #5111N-4 Batteries are...
Page 6
Corrective ction Aperture dial is Rotate Aperture dial between positions Lamp Burned Replace the lamp with No light output ADC part #5112N-4 Wrong Lamp Replace the lamp with Installed ADC part #5112N-4 Batteries are depleted Replace with fresh “AA” batteries...
To Obtain Warranty Service: Send item(s) postage paid to ADC®, Attn: Repair Dept., 55 Commerce Dr., Hauppauge, NY 11788. Please include your name and address, phone no., proof of purchase, and a brief note ex- plaining the problem.