Otoscope and Accessories Troubleshooting Guide
Trouble Area
Possible Cause
Lamp burned out
Wrong lamp Installed
No light output
Batteries are depleted
Lamp not fully inserted
Instrument head
Thread/connection
will not connect
Trouble Area
Possible Cause
Specula will not
attach properly
Loose or misaligned
magnifying lens
Poor or obstructed
view through lens
Lens is damaged
Dim light
output
Visible corrosion on
instrument or accessories
excessive moisture
Ophthalmoscope Troubleshooting Guide
Trouble Area
Possible Cause
between positions
No light
output
Instrument
head
will not connect
Thread/connection
Spot is not
The aperture dial
centered
Dim light
output
Aperture or lens wheel
Dirt in or damage
will not turn or does so
with difficulty
Visible corrosion
on instrument
excessive moisture
Environment
Transportation / Storage:
Environmental Operating Range:
Max. Operating:
Corrective Action
Replace the lamp
Replace the lamp
Replace with fresh
AA batteries
Re-insert lamp
Service
is damaged
required
Corrective Action
Specula mount
Service
is damaged
required
Lens mount
Service
is damaged
required
Lens is dirty
Clean with lint-free cloth
Service required
Lamp near end
Replace the lamp
of useful life
Batteries are
Replace with fresh
depleted
AA batteries
Damage from
Service
required
Corrective Action
Aperture dial is
Rotate aperture
dial
Lamp burned
Replace the lamp
out
Wrong lamp
Replace the lamp
installed
Batteries are
Replace with fresh
depleted
AA batteries
Lamp not fully
Re-insert
inserted
lamp
Service
is damaged
required
Move aperture dial to
is not centered
the full detent position
Lamp near end
Replace the lamp
of useful life
Batteries are
Replace with fresh
depleted
AA batteries
Service
to
required
mechanism
Damage from
Service
required
-4°F to 120.2°F (-20°C to 49°C)
Relative Humidity: 95 %
500 hPa – 1060 hPa, Altitude
50°F to 120.2°F (10°C to 49°C)
RH 95 % (Max)
500 hPa – 1060 hPa, Altitude
95°F (35°C)
Pocket Diagnostic Instruments
Instructions for Use
LIMITED WARRANTY
ADC
®
warrants its products against defects in materials and workmanship under
normal use and service as follows:
1. Warranty service extends to the original retail purchaser only and
commences with the date of delivery.
2. The instruments are warranted for two years. The LED lamps are warranted for life.
What Is Covered:
Repair, or replacement of parts, and labor.
What Is Not Covered:
Transportation charges to ADC. Batteries where supplied. Damages caused by
abuse, misuse, accident, or negligence. Incidental, special, or consequential
damages. Some states do not allow the exclusion or limitation of incidental,
special, or consequential damages, so this limitation may not apply to you.
To Obtain Warranty Service:
Devices should be contained in their transport/storage cases when being
returned to ADC to ensure that they are not damaged during transport and
handling. Do not ship contaminated devices or storage cases to ADC's facility.
Send item(s) postage paid to ADC, Attn: Repair Dept., 55 Commerce Dr.,
Hauppauge, NY 11788. Please include your name and address, phone no.,
proof of purchase, and a brief note explaining the problem.
Implied Warranty:
Any implied warranty shall be limited in duration to the terms of this
warranty and in no case beyond the original selling price (except where
prohibited by law). This warranty gives you specific legal rights and
you may have other rights which vary from state to state.
To register your product,
visit us at
www.adctoday.com
and follow the links
ADC
55 Commerce Drive
Hauppauge, NY 11788
U.S.A.
SC Cattus SRL
Aleea Carpati nr. 53 Sc. B Ap. 64
Târgu-Mures, România, EU
IB p/n 9362-00 rev 10 (09/23/2021)
Inspected, assembled and
packaged in the U.S.A.
Components made in Pakistan
tel: 631-273-9600
toll free: 1-800-232-2670
fax: 631-273-9659
www.adctoday.com
info@adctoday.com
MD
Printed in the U.S.A.

Advertisement

loading

Summary of Contents for ADC Diagnostix 5110N Series

  • Page 1 AA batteries Devices should be contained in their transport/storage cases when being Lamp not fully Re-insert returned to ADC to ensure that they are not damaged during transport and Instrument inserted lamp handling. Do not ship contaminated devices or storage cases to ADC’s facility.
  • Page 2 Thank you for choosing an ADC diagnostic set. We’re proud of the care and Attaching the Disposable Specula quality that goes into the manufacture of each and every diagnostic instrument Push the flanged end of the desired speculum onto the that bears our name.

This manual is also suitable for:

Diagnostix 5110e seriesDiagnostix 5111n seriesDiagnostix 5112n series5110n5110egr5110ngr ... Show all