12. Insert Autosoft 90: Position AutoSoft
90 over the insertion site. Simultaneously
press the round indentations on both
sides of the insertion device to insert
AutoSoft 90.
13 . Push gently on the center of the inserter to
secure the adhesive on the skin .
14 . Remove the inserter/introducer needle by
grasping the center and pulling it gently
back. Massage the tape thoroughly
onto the skin. Fill the empty space in the
cannula with the following amount of
insulin:
U-100 INSULIN:
6 mm: 0.3 units (0.003 ml)
9 mm: 0.5 units (0.005 ml)
15 . Push the lid back in place until you hear
a "click . "
DISCONNECT
i .
AutoSoft 90 allows you to temporarily
disconnect your pump without having
to change the infusion set. Hold the
adhesive in place by placing a finger just
in front of the cannula housing, and gently
squeeze the sides of AutoSoft 90. Pull the
connector needle straight out from the
cannula housing .
ii . Insert the disconnect cover into the
cannula housing until you hear a "click".
RECONNECT
iii . Prime the infusion set until medication
dispenses from the needle.
When priming the infusion set,
CAUTION:
hold AutoSoft 90 with the needle pointing
down.
iv . Place a finger on the adhesive tape just in
front of the cannula housing.
v . Push the connector needle straight in until
you hear a "click".
ENGLISH
7