Care and Maintenance

Preparation Prior to Cleaning

Disassembly
Disassemble device if applicable. Remove all instruments and
attachments from the power tool and remove battery from
casing or handpiece.
To clean the batteries and the charger, wipe them off with a
clean, soft and lint-free cloth dampened with disinfectant or
deionized water (Fig. 1 and 2).
Important: Do not use solvents to disinfect the batteries.
Battery poles must not contact water or solvents: danger of
short circuiting!
Return batteries to charger (05.001.204) after each use
(Fig. 3).
Never immerse the handpiece, lid or attachments in aqueous
solutions or in an ultrasonic bath as this could decrease the
service life of the system.
Handpieces and attachments may be processed using
a) manual cleaning and/or
b) automated cleaning with manual pre-cleaning.
Note: Clean all movable parts in opened position.
30
Synthes
Colibri II
Instructions for Use
Figure 1
Figure 2
Figure 3