6. Commissioning, operation
6.9.2.1 ZERO submenu item
EN
Menu item
AUTO-ZERO
INTERVAL

6.9.2.2 RS232/USB submenu item

For information on this menu item, please refer to chapter 7 "Interface".

6.9.2.3 DISPLAY submenu item

Adjustment of the brightness of the display can be set from 0 ... 100 %.
1. Set the value range 0 ... 100 % with the [▲] / [▼] buttons.
2. Confirm with [SELECT].
28
ZERO
-------------------------
AUTO-ZERO
INTERVAL
Meaning
Switch zero point correction on or off
OFF: Zero point correction can be carried out manually via the [ZERO]
button
ON: Zero point correction is carried out automatically
1. Activate or deactivate with the [▲] / [▼] buttons
2. Confirm with [SELECT].
Zero point correction is carried out in adjustable time intervals
The range is 1 ... 60 minutes
1. Set the value range 0 ... 100 % with the [▲] / [▼] buttons.
2. Confirm with [SELECT].
DISPLAY
-------------------------
BRIGHTNESS:
OFF
60min
100%
WIKA operating instructions, model CPC2000