4 Important Information
These electronic balances are precision instruments. It is designed for use in a self-
controlled electromagnetic environment. That means that in such an environment
radio equipment, such as mobile phones, must not be used in direct vicinity. In case
strong electromagnetic fields should arise, deviations of the indication are possible.
Avoid disturbing environmental conditions such as currents and vibrations as well as
a rapid change of the ambient temperature. If necessary readjust the balance after a
change of temperature.
Avoid high humidity, vapours and dust as the balance is not hermetically closed. Do
not let the balance come into direct contact with liquids as these could enter the
measuring cell. Therefore only clean the balance with a dry or damp cloth. Do not
use solvents as varnished or plastic parts could be damaged.
A warming up time of a few minutes after switching the balance on will stabilise the
measuring values. Place the object carefully on the weighing pan. Avoid any constant
load on the weighing pan even when out of use. Avoid shocks, thrusts and extreme
impairments by all means. Remove spillages immediately.
When the programme cycle is disturbed switch the balance off and then on again.
Start the weighing operation again.
Never use the balance in rooms where there is a danger of explosion. The serial
version is not explosion-proof.
Opening of the balance causes invalidity of warranty.
Check the balance daily with the supplied check weight.
If necessary adjust the balance. The warranty is nullified if the balance is opened.

5 Minor breakdowns

Fault
The weight display does
not light up
The weight display
changes constantly
The display indicates "L0"
The weighing result is
obviously incorrect
If any other error messages appear turn off the balance and on again. When the error
message still appear please contact the manufacturer.
14
Possible Cause
• The balance is not switched on
• The balance is not connected to the mains (mains
cable not plugged in or defective)
• The is a power failure
• Draught/air currents
• Vibrations of the table/floor
• The weighing pan is touching a foreign body
• The battery is dead. Change battery.
• The adjusting is no longer correct
• The balance was not tared
• The room temperature is fluctuating noticeably.
466-BA-defsi-0016

Advertisement

Table of Contents
loading

This manual is also suitable for:

466-41466-45