Installing/mounting

3.2 Sensor installation

Loss of type of protection
Risk of explosion. Damage to device if the enclosure is open or not properly closed. The type
of protection specified on the nameplate or in Technical data (Page 57) is no longer
guaranteed.
● Make sure that the device is securely closed.
3.2
Sensor installation
3.2.1

Determining a location

Electromagnetic fields
Do not install the flowmeter in the vicinity of strong electromagnetic fields, for example near
motors, variable frequency drives, transformers etc.
Upstream / downstream
● No pipe run requirements, that is straight inlet/outlet sections, are necessary.
● Avoid long drop lines downstream from the sensor to prevent process media separation
causing air / vapor bubbles in the tube (min. back pressure: 0.2 bar).
● Avoid installing the flowmeter immediately upstream of a free discharge in a drop line.