Error Messages
• Measurement value
blinks.
• "bt out" will not be shown
when the battery is
completely exhausted.
• Battery indicator is
flashing.
• Display for about 3
seconds and the Power
is switched off.
E8
Cause
• The SpO
value is below
2
the lower limit (70% SpO
of the measurement
range. The measured
value will still be
displayed, but its accuracy
cannot be guaranteed.
• Battery Out
• Battery power is nearly
empty.
• Other trouble occurred in
the measurement circuit
of the instrument.
Solution
(This message indicates that the
measured value is outside the
)
2
accuracy guaranteed range.)
• Replace the battery with new one.
• In this case, the measurement data
is not left if the battery is removed.
• Battery will be empty in about 30
seconds. Remove battery and
replace with new one.
• Turn the Power Off, and turn ON
again.
• If these message reappear even
though the above points have been
checked, the instrument may be
malfunctioning. Contact to the
nearest distributor or authorized
service facilities.