OXYGEN SATURATION MONITOR -300 -300 -300 PULSOX ® ® Instruction Manual This instrument must be used according to the instructions of the doctor. Get other manuals https://www.bkmanuals.com...
Safety Symbols Warnings and precautions noted in this manual and the instrument are indicated by the following markings, designed to prevent accidents caused by erroneous handling of the equipment. This indicates text consisting of a warning or precaution relating to safety. Please read the text carefully and use the equipment safely.
Safety Precautions To ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises. WARNING (Failure to adhere to the following points may result in death or serious injury.) The instrument is designed for measurement of the ox-...
Page 4
Do not use probes other than those specified by KONICA MINOLTA SENSING, INC. Use of alterna- tive probes may cause the probe to overheat, result- ing in burns.
Contents Safety Symbols Safety Precautions .............1 Foreword ................4 Safety Notes ................4 Package Contents ..............5 About Probe ................5 System Diagrams and Accessories ........6 Names of Parts ..............7 Notes on Use ................. 8 Notes on Storage ..............10 Disposal Method ..............11 Preparations ..............12 1 ) Attaching the Wrist Band ..........
Do not use it for any other purposes, such as warning warning of sleep apnea and breathing abnormalities. The Oxygen Saturation Monitor PULSOX-300 is designed to measure oxygen saturation (SpO value) in arterial blood and pulse rate by attaching a probe to an adult fi...
Package Contents Before using the instrument, make sure that the following items are present. 1) PULSOX-300 main body × 1 2) Wristband WB-300 × 1 3) AAA-size alkaline battery × 1 4) Instruction manuals • PULSOX-300 instruction manual × 1 •...
System Diagrams and Accessories The instrument consists of the following items. Standard accessories Optional accessories Optional accessories Finger Clip Probe SR-5C Spot Check Probe SP-5C Extension cable EC-300 ���������� Monitor Probe LM-5C Personal Probe SD-5C Standard accessories Wristband WB-300 AAA-size alkaline battery Get other manuals https://www.bkmanuals.com...
Names of Parts Power button Used to start/end measurement. Display (LED) Measurement results and messages are displayed. Probe connector Connect the probe to this connector. Battery cover The cover can be opened by turning it approximately 90 degrees. Wristband Used to set the instrument on your wrist.
Notes on Use • This instrument should be used under the following operat- ing conditions. Temperature of 0 to 40°C (32 to 104°F), relative humidity • of 30 to 85%, with no condensation. Atmospheric pressure of 700 to 1060 hPa (altitude of •...
Page 11
When influenced by abnormal hemoglobin like carbon • monoxide hemoglobin (HbCO) and methemoglobin When pigments such as cardiogreen, intravascular dyes • and indocyanine greenare present in the blood When fi nger nails are polished When fi nger nails are polished •...
Notes on Storage • This instrument should be stored under the following storage conditions. Temperature of –10 to 60°C (14 to 140°F), relative humid- • ity of 10 to 95%, with no condensation. Atmospheric pressure of 700 to 1060 hPa (altitude of –400 •...
Disposal Method • When disposing of the used battery, insulate the polarity ter- minals with insulating tape etc. If the terminals of the battery come into contact with metal objects, heat generation, explo- sion or fi re may result. • Make sure that the battery is either disposed of or recycled correctly in accordance with local laws and regulations.
Preparations Before starting measurements using this instrument, follow the procedure given below to ensure that measurements are stable. 1) Attaching the Wrist Band Attach the Wrist Band WB-300 (supplied as a standard accessory) to the instrument. Since the WB-300 is made of elastic material, it allows for fl...
◆ D o n o t c o n n e c t i t i n w i t h *Connecting the extension cable excessive force. Doing so may EC-300 between the PULSOX-300 damage the connector. and the probe extends the probe cable ◆...
3) Installing the battery Procedure Turn the battery cover clockwise approximately 90 degrees to open the cover. ◆ W h e n t u r n i n g t h e c o v e r, k e e p t h e f l a t part of your finger on the coverʼs catch.
fi nger of an adult patient. Press the Power button. ● The Power is ON, all the elements on the display are displayed for about PULSOX-300 2 seconds, and then the display changes to the measurement display. ◆ Until the measurement...
2) Reading the display When the instrument is turned ON, all the elements of the LED will light up. The pulse level, pulse rate and SpO The pulse level, pulse rate and SpO value are displayed value are displayed during measurement. Pulse rate (P.R.) Battery indicate If an error occurs, an...
3) Ending the Measurement Press the Power button, the measurement is finished and the Power is off. PULSOX-300 About Auto Power-Off function The instrument has an auto power-off function that operates independently. If no probe is connected to the PULSOX ("C"...
Check Points before repairing Before taking any actions, check the following points fi rst. ◆ If the abnormality continues to reappear, the instrument might be faulty, so contact the nearest distributor or authorized service be faulty, so contact the nearest distributor or authorized service facility.
■ Dimensions (W × H × D) 68 × 58 × 15 mm ■ Weight 56g (including battery/excluding wrist band and probes) ■ Usable life 6 years [Verifi ed by KONICA MINOLTA SENSING, INC. (based on own data)] Get other manuals https://www.bkmanuals.com...
Page 26
Probes (SR-5C, SP-5C, LM-5C, SD-5C) Probes (SR-5C, SP-5C, LM-5C, SD-5C) ■ LED-SPD probe ■ Type • Finger Clip Probe SR-5C • Spot Check Probe SP-5C • • Monitor Probe LM-5C Monitor Probe LM-5C • Personal Probe SD-5C It can be used with one of Extension Cable EC-300 ✻...
Appendix Calculation Method for Displayed Values <Pulse level meter> The pulse level meter displays the transmitted light amount ratio calculated by the following equation. The pulse level is indicated in eight levels (0 to 15%, full scale if over 10%). Transmitted light amount that varies due to the pulse Pulse level (%) = ------------------------------------------------------------------------------------------------------------------ x 100 Transmitted light amount that does not vary due to the pulse...
EMC Guidance Guidance and manufactureʼs declaration - electromagnetic emissions The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSOX-300 should assure that it is used in such anenvironment. Emissions test Compliance Electriomagnetic environment - guidance RF emissions The Model PULSOX-300 uses RF energy only for its internal function. Therefore its RF emissions are CISPR 11 Group 1 very low and are not likely to cause any interference in nearby electromagnetic equipment. RF emissions The Model PULSOX-300 is suitable for use in all Class B establishments, including domestic establishments CISPR 11 and those directly connected to the public Harmonic emissions low-voltage power supply network that supplies Not applicable IEC61000-3-2 buildings used for domestic purposes. Voltage fluctuations/ flicker emissions Not applicable IEC61000-3-3 Guidance and manufactureʼs declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSOX-300 should assure that it is used in such anenvironment. Immunity test IEC 60601 Compliance Electromagnetic environment - Test level...
Page 29
Guidance and manufactureʼs declaration - electromagnetic immunity The Oxygen Saturation Monitor Model PULSOX-300 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model PULSOX-300 should assure that it is used in such anenvironment. Immunity test IEC 60601 Compliance Electromagnetic environment - Test level level guidance Conducted RF 3 Vrms 3 Vrms Portable and mobile RF communications IEC61000-4-6 equipment should be used no closer to 150kHz to 80MHz any part of the PULSOX-300, than the Radiated RF 3 V/m 3 V/m recommended separation distance IEC61000-4-3 80MHz to 2.5GHz calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance 80MHz to 800MHz ������������ � ��� 800MHz to 2.5GHz ������������ � where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance inmeters (m).
Measurement Principle This instrument measures the oxygen saturation (SpO ) in arte- rial blood and pulse rate continuously in a non-invasive method by applying optical principles. by applying optical principles. With this instrument, SpO is defi ned by the following equation. where C (Hb): Concentration of reduced hemoglobin...
Relation between Oxygen Saturation and Partial Pressure The relation between oxygen saturation (SpO ) and oxygen partial pressure (PaO ) is shown in the graph below. SpO is the oxygen saturation as measured by pulse oximeters. oxygen saturation as measured by pulse oximeters. Oxygen Saturation vs.