AEMC 6416 Service/Calibration Manual page 13

Clamp-on ground resistance tester
Hide thumbs Also See for 6416:
Table of Contents

Advertisement

MODEL 6416/6417
User Current Calibration
1.3.9 Engage the jaws of the ground clamp around the current carrying conductor.
1.3.10 Navigate to the "CAL. I" menu.
1.3.11 Press and hold the ► key until the "PrES. rt." and "100.0 mA SEt" messages
appear and start alternately blinking (press should last approximately 3
seconds).
1.3.12 Run a known current between 50 and 150 mA (60 Hz) through the current
carrying conductor.
1.3.13 Enter the current value using the ▲ and ▼ keys and press the ► key to start
the low current section of the calibration ("run CAL.I" will be displayed for
approximately 10 seconds).
1.3.14 When the "PrES. rt." and "10.00 A SEt" messages appear and start
alternately blinking, run a known current between 9 and 10.5 A (60 Hz)
through the current carrying conductor.
1.3.15 Enter the current value using the ▲ and ▼ keys and press the ► key to start
the high current section of the calibration ("run CAL.I" will be displayed for
approximately 10 seconds).
1.3.16 "End CAL.I PASS" or "End CAL.I FAIL" will be displayed at the end of the
calibration. Press the ► key to save the new user calibration and return to the
"CAL.R" menu.
1.3.17 Retry the calibration if failed.
Note: if an adjustment operation fails (FAIL message), check that there is nothing interfering
with the closing of the clamp, then repeat the operation. If the problem persists, the clamp
must be sent back for repair to the manufacturer.
v1 – 11/23/15
SERVICE/CALIBRATION MANUAL
- 13 -
CAT# 2141.01/2141.02