Make unlock card touch response area of the equipment;
C area: shows duration and vacuum of handpiece 4.

5.5 Ultrasonic Frequency

Picture 5.5-1
After you change the power supply, ultrasonic handpiece, or find energy increase or decrease, you
shall adjust the frequency.
1) Press
key for 3s(Picture 5.5-1); then system shows Picture 5.5-2;
2) Connect the handpiece needing check with the left connector, not the right connector (Picture
5.5-3); then relative picture turns yellow(Picture 5.5-2);
3) Bipolar cavitation of handpiece 1 is adjusted together;
4) Bipolar cavitation of handpiece 2 is adjusted separately; choose the left one or right one by
touching picture ;
5) Leave some water on the cavitation part, press "O" key, find the best frequency of water
shaking by "
be consistent;
6) After adjustment, press "O" key, press
failure and press
7) Press
to go back to main interface;

5.6 IC Card

The issued time is saved in the IC card, while, the equipment doesn't save it.
So equipment has no function of loss report. But the equipment could manage countless persons.
13//22
Picture 5.5-3
" key (Picture 5.5-3) :the two parts' water shaking of handpiece 1 shall
, adjust it again;
Picture 5.5-2
key;
means success and
means
All rights reserved