Memory Clearing, Sanitization and/or Removal Procedures
Table 6-2
Description and
purpose
Size
Memory clearing
Memory
sanitization
Memory removal
Write protecting
Memory
validation
Remarks
Keysight N991xA/N992xA/N993xA Security Features and Volatility Documentation
4 GB Internal microSD Card
Stores Operating System, Application Firmware and some User Data. Only the
User Data is accessible by the user
4 GB
The following process completely clears all user accessible data on the
FieldFox instrument. This process requires firmware revision A.03.02 or later.
Press the following keys:
1. System(7)
2. Service Diagnostics
3. Advanced
4. Erase User Data
5. Confirm Erase
Wait for the instrument to reboot and begin normal operation.
The process will completely erase the UserData on the 4GB Internal microSD
Card and reboot the instrument. When the reboot is completed, the
instrument will be ready for normal operation.
The following process completely clears all user accessible memory on the
FieldFox instrument. This process requires firmware revision A.03.02 or later.
Press the following keys:
1. System(7)
2. Service Diagnostics
3. Advanced
4. Erase User Data
5. Confirm Erase
Wait for the instrument to reboot and begin normal operation.
The process will completely erase the Data on the 4GB Internal microSD Card
and reboot the instrument. When the reboot is completed, the instrument will
be ready for normal operation.
This memory can not be removed without damaging the instrument
N/A
N/A
4GB microSD is only used in FieldFox model N9913A, N9914A, N9915A,
N9916A, N9917A, N9918A, N9925A, N9926A, N9927A, N9928A, N9935A,
N9936A, N9937A, N9938A
UserData on internal flash (64 MB flash memory) is not available when
microSD flash memory is used.