Microstream
Instructions for use:
1.
Place the monitor outside the MRI suite. There must be
a hole in the wall of the suite (approximately 10 cm di-
ameter).
2.
Connect the FilterLine
ing the female luer connector securely but gently into
the Microstream
3.
Guide the FilterLine
the MRI suite.
4.
Attach the FilterLine
ed Microstream CO
3
patient
sampling line to the luer connector on the FilterLine XL.
5.
Ensure that the tubing is not twisted or crimped.
6.
Attach the CO
in its Directions for Use.
7.
Following connection of the CO
that CO
2
1
2
3
®
®
®
input connector
®
XL through the hole in the wall of
®
XL to an intubated or non-intubat-
sampling line as appropriate for the
2
, by attaching the male luer at the end of the
sampling line to the patient as described
2
values appear on the monitor display.
EtCO
Sampling
2
1
XL
to the monitor by twist-
2
sampling line, check
2
.
3