IEC 60601-1. If the CA-200F corneal analyser is installed in rooms for medical use and is not connected to a computer, there is no need to use an insulated transformer. If a printer is connected directly to the CA-200F corneal analyser, it must be powered by an insulated transformer in accordance with IEC 60601-1.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 INTENDED USE Use: CA-200F is a corneal analyser with integrated pupillograph. The instrument acquires images of the cornea and analyses its topography. The software selects the image with the best focus out of a sequence of images.
This instrument also allows you to simulate contact lenses and view the 3D map to analyze the wavefront corneal aberrations. Notes: This manual describes the TOPCON CA-200F corneal analyser, including the functions, basic operations, instrument cleaning, and instrument storage. For best use of the instrument, carefully read the instructions provided.
EMC table Emission aspects The CA-200F device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is used in the said environment. Electromagnetic environment – guidance...
Page 6
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Immunity aspects The CA-200F device is intended for use in the electromagnetic environment specified below. The customer or the user of the CA-200F device should assure that it is used in the said environment. Electromagnetic environment – EN 60601-1-2...
Page 7
Radio frequency immunity aspects The CA-200F device is intended for use in the electromagnetic environment specified below. The customer or the user of the CA-200F device should assure that it is used in the said environment. Electromagnetic environment – EN 60601-1-2 test...
Page 8
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Symbols Description publications CLASS II DEVICE IN ACCORDANCE WITH 60601-1 IEC 60417-5840 APPLIED PART TYPE B IEC 60417-5172 INSULATION CLASS II DEVICE (DOUBLE INSULATION) PRODUCT IN ACCORDANCE WITH DIRECTIVE 93/42/EEC IEC 60417-5032 ALTERNATING CURRENT...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 WARNINGS FOR SAFE USE In order to safely use the instrument and prevent risks to the operator and other persons as well as damage to the device, the instruction manual provides a description of the safety warning labels and plates on the instrument body.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 USE AND MAINTENANCE As the CA-200F Corneal Analyser is an electronic instrument for medical purposes, it must be used by expert and qualified staff. MAINTENANCE To ensure the safety and performance of the equipment, it is advisable not to perform operations different from those indicated below.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 WARNING LABELS AND PLATES In order to safely use the instrument and prevent risks to the operator and other persons as well as damage to the device, the instruction manual provides a description of safety warning labels and plates on the instrument body.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 COMPONENTS MAIN COMPONENT PARTS Placido disc Power connector Display USB port ON/OFF LAN port button Acquisition button Joystick Function buttons...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 CA-200F STANDARD ACCESSORIES The following accessories are included in the package: POWER CABLE TRANSFORMER CALIBRATION TEST INSTRUMENT INSTRUCTION AND ASSEMBLY MANUAL DECLARATION OF CONFORMITY TOUCH SCREEN PEN USB EXTENSION CABLE...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Installation without network connection Transformer Power socket Connection to an existing network Network Hub LAN connection (paired cable) Transformer Power socket...
When the device is turned on, the software is automatically loaded. CA-200F can operate in two different ways: exchanging data with a PC or stand-alone. Described below are the acquisition procedures for operation connected to a remote PC as well as in stand- alone mode.
Page 19
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-1 You can insert a new patient or select an existing patient from the database of the PC connected to the analyser. To insert a new patient, fill in the patient details fields using the on-screen keyboard. If you need to insert special characters, keep pressing the letter buttons: a panel with the choice of the possible characters for the fig.
Page 20
CA-200F Corneal Analyser - Rev. 7 16/01/2012 If the communication between the PC and the device is correctly established, the patients that match fig. 1-4 the criteria entered will be displayed in the dedicated section ( ) . Select the patient and press the “Open”...
Page 21
CA-200F Corneal Analyser - Rev. 7 16/01/2012 The following buttons are also available in the Patient Details window: fig. 1-6 Local List: opens a list with locally stored items ( ). This button is only active when there are locally stored items available.
Page 22
CA-200F Corneal Analyser - Rev. 7 16/01/2012 ACQUISITION ENVIRONMENT The acquisition environment is divided into three sections: Topography, Pupillometry and Fluorescein. fig. 1-8 Each time you access acquisition, the instrument is set to topography ( fig. 1-8 The following buttons are available in the acquisition window: R and L: indicates the eye being acquired, highlighted in yellow.
Page 23
CA-200F Corneal Analyser - Rev. 7 16/01/2012 The images are numbered progressively and each of them is associated with the eye to which they refer and the type of acquisition. For topography and fluorescein, you can tap on a preview image to select or deselect it. The selected images are displayed normally, while those not selected are dark.
Page 24
CA-200F Corneal Analyser - Rev. 7 16/01/2012 TOPOGRAPHY Backlighting of the Placido disc is automatically activated when the acquisition environment is accessed. If the analyser is not used for a few minutes, the cone turns off; to switch it on again, just push the button on fig.
Page 25
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-14 Acquisition procedure Align the live image in the center and focus it, then push the joystick button to start acquisition. You can stop acquisition by pushing the joystick button. Otherwise the acquisition is automatically interrupted when the sliding bar reaches the end.
Page 26
CA-200F Corneal Analyser - Rev. 7 16/01/2012 SETTINGS In the acquisition window press the “Settings” button ( fig. 1-16 The first time the program is started, the default language set is English and keyboard layout is “QWERTY”. To change the language setting, select the desired language from those that appear when tapping on the ...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-17 ACQUISITION IN STAND-ALONE MODE To use the instrument in stand-alone mode, the PC Link software module must be deactivated. fig. 1-18 The acquisition environment is as shown in fig. 1-18 Tapping on the “New Patient” button, a window opens where you can insert the patient details (...
Page 28
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-19 to close the window. The “New Patient” button Insert the patient data and tap on the button changes to “Close” ( fig. 1-20 Tapping on “Close” the patient data and the gallery are deleted. The “Close” button changes back to “New Patient“...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 TOPOGRAPHIC MAP In the acquisition window press “Processing” to access the topographic map environment ( fig. 1-21 fig. 1-21 Tap on the “R” or “L” button to view the map of the right eye (R) or the left eye (L).
Page 30
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Profile Press the "Profile" button to view the curvature profile along the most curved and the flattest meridian (red and blue). fig. 1-24 The difference is shown in green ( fig. 1-24 Tap on the “>>” button to open the panel ( fig.
Page 31
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Keratometry Tap on the ”K” button to view the keratometric data ( fig. 1-27 Sim K Ophthalmometric reading simulation Meridians Keratometric data at 3 mm, 5 mm and 7 mm Emimeridians Keratometric and emimeridian data at 3 mm, 5 mm and 7 mm fig.
Page 32
Keratoconus Tap on the “KC” button to open keratoconus screening with the following information fig. 1-29 AK: Apical curvature. Represents the cornea power at its apex AGC: apical curvature gradient. Represents the mean variation by unit of length of the cornea power taking the apical power as reference SI: difference between the average power of two circular areas centered on the vertical axis of the rulers and positioned in the lower and upper hemisphere of the cornea,...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-31 Acquisition Tap on "Acquisition" to return to the acquisition window. Modules fig. 1-32 Press the "Modules" button to open the software modules ( Fluorescein Pupillometry and Zernike (if active in the instrument)
Page 34
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-33 Scales fig. 1-34 Select a unit of measure: Diopters Millimeters Asphericity Select an asphericity unit of measure:...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 FLUORESCEIN fig. 1-35 The fluorescein module ( ) allows assessing the physical condition of the cornea and positioning and dragging the contact lenses. fig. 1-35 The pictures and movies acquired are viewable in the gallery.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-36 Functions Settings Opens the fluorescein options window. Report Save the report as a PDF file to a USB unit or print it on a connected USB printer. Acquisition Returns to the acquisition window.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 PUPILLOMETRY (optional module) The pupillometry module allows viewing and analyzing the dynamic and static pupillometry (images of the pupil acquired in controlled light conditions). fig. 1-39 The software by default accesses dynamic pupillometry, if acquired ( fig.
Page 38
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Photopic, Scotopic, Mesopic Tap on the “P”,”S”, “M” buttons to view the images of the pupil acquired in fig. photopic, scotopic and mesopic conditions with the following information ( 1-42 Average value and standard deviation of the average pupil diameter in the...
Page 39
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Decentralization fig. 1-44 The green box identifies the pupil center coordinates with respect to the fixation point. The red lines show how these coordinates vary during dynamic pupillometry acquisition. On the left-hand side are the minimum and maximum values of the pupil diameter acquired in dynamic pupillometry and the pupil center, Cartesian and polar coordinates.
Page 40
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Considering that dynamic pupillometry consists of acquisition of a sequence of images as the light conditions change from scotopic to photopic and back to scotopic and that the scotopic (red band) and photopic (green band) acquisition times can be set in “Settings”...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 PUPILLOMETRY SETTINGS fig. 1-47 fig. 1-48 fig. 1-47 fig. 1-48 Select one or more options to customize the display of the pupil images ( ) with: Grid Ruler Ring Center: the center of the pupil will be displayed in blue and the fixation point in red...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 ZERNIKE (included with pupillometry module) The Zernike module allows full display of the wavefront aberrations generated by the front surface of the cornea. The results of the Zernike analysis are displayed by means of numerical indices and graphic fig.
Page 43
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-50 Functions Pupil The “Pupil” button opens a panel ( fig. 1-51 ) where you can select the pupil diameter (in a range between 2 mm and 7.5 mm) to see how the aberrations change as the pupil diameter changes.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 LENSES fig. 1-52 The lenses module ( ) simulates contact lens positioning. fig. 1-52 Tap on the "R" or "L" button to view the lens in the right or left eye. Selecting a lens fig.
Page 45
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 1-54 fig. 1-55 fig. 1-56 fig. 1-57 Profile fig. 1-58 Tap on the "Profile" button to view the lachrymal film profile under the lens selected ( fig. 1-58 Tap on the “>>” button to open the panel ( fig.
Page 46
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Lens indices and options fig. 1-60 You can select the following buttons ( Ref: Refraction K/L:Keratometry/Limbus T/D:Tilt/Decentration Pref: Lens preferences fig. 1-60 Refraction fig. 1-61 Shows the refraction values ( Tap on “>>” to set the refraction values ( fig.
Page 47
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Tilt/Decentration fig. 1-64 Tilt fig. 1-64 Displays the four buttons in lens simulation ( ). Press the buttons to set the different tilt values. The "Reset" button resets the initial condition. Decentration Drags the lens to the desired position. The "Reset" button resets the initial lens position.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 The "Reset" button deletes all the lens preferences. The "Close" button closes the window. Functions Settings Opens the lens options window. Report Save the report as a PDF file to a USB unit or print it on a connected USB printer.
Page 49
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Lenses fig. 1-67 fig. 1-67 Set the apical clearance using the up/down arrows (...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 TORIC IOL (optional module) NOTE: All the information contained in this paragraph are referred to IOL version 2.0.0.5. The Toric IOL software module is the tool to calculate Oculentis toric intraocular lenses. fig. 1-68 fig.
Page 51
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Surgeon Information Select a surgeon from the list. fig. 1-70 Tap on "Details" for the surgeon details ( Tap on "Add new" to add a new surgeon. fig. 1-70 Tap on "Close" to close the surgeon details.
Page 52
CA-200F Corneal Analyser - Rev. 7 16/01/2012 The following information is required: Biometry Procedure Axial Length IOL Spherical Power Surgically Induced Astigmatism Incision Location fig. 1-72 Tap on "Calculate IOL" to view the result window with the pre-op and post-op information ( fig.
- Decompress the upgrade package in the root (main folder) of an empty and FAT32 formatted USB pen-drive - Insert the USB pen-drive in the CA-200F device - Tap on the “Upgrade” button The system will be restarted for the first time. Disconnect the USB pen, otherwise the system will not be started.
Page 54
CA-200F Corneal Analyser - Rev. 7 16/01/2012 Terminate Press "Terminate" to close the CA-200F application and return to the Windows desktop. Network Settings Press “Network Settings” ( fig. 1-76 Use the “Search” button to start the search for the IP address for connection to the PC.
CA-200F Corneal Analyser - Rev. 7 16/01/2012 MODULE UPDATE From the “Admin Tools” panel, tap on the "Module Update" button ( fig. 1-78 Here you will find the sections for activation of the optional and software modules. fig. 1-78 Optional Module...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 2. CA-200F PC-SOFTWARE SOFTWARE INSTALLATION See the “CA-Series Installation Manual” included in the Setup CD. CA-200F PC SOFTWARE SETTINGS For details, refer to the “Corneal Analyser Software Manual”. To open the settings window, in the main “Corneal Analyser Software” window ( ) tap on “File”...
Page 57
CA-200F Corneal Analyser - Rev. 7 16/01/2012 fig. 2-2 To select the acquisition device, tap on “Corneal Analyser”, select CA-200F as instrument and press “Ok” to close the window and save the changes. Tap on “Cancel” to close the window without saving the...
CA-200F Corneal Analyser - Rev. 7 16/01/2012 3. TROUBLESHOOTING Problem Solution Impossible to communicate with the PC Check that the network has been installed as described in the “CA-Series Installation Manual” (This may occur while searching the patient list and/or when sending patient data to the PC)
If the instrument is dirty, clean the surface with a dry cloth. If there are permanent stains on the surface of the Placido disc, please contact Topcon Support for replacement. Before using the chinrest on another patient, clean and disinfect the supports that come into contact with the forehead and chin with neutral detergents.
Page 62
Right alignment Wrong alignment fig. 6-2 From the CA-200F software, select the "Settings" button and then the “Calibration Check" function ( fig. fig. 6-3 Following the instructions given at the bottom of the window, acquire the sphere eight times with a radius of 8 mm.