PM100 Bump Test and Calibration Station
5.3.1 Station Tab
The Station tab is displayed when the software opens (as shown above). It shows information about the Bump-
Calibration Station, information about the installed gas cylinder, configuration of the operation of the Bump-
Calibration Station, and the Bump-Calibration Station firmware upgrade interface.
NOTE: No changes will be saved to the Bump-Calibration Station until the Write button is pressed (see 5.3.1.1).
5.3.1.1
Station Info
Serial Number – Cannot be changed. It uniquely identifies the Bump-Calibration Station unit.
Version – The firmware version installed on the Bump-Calibration Station.
Auto Power-off – Determines the idle time before the Bump-Calibration Station automatically switches to sleep
mode. Can be set from 1-3600 seconds. Default is 300 seconds.
Maximum Bump Time – Determines the maximum time allowed for a Bump Test. If the PM
device fails to
100
R
R
properly respond before the time expires, the test will stop, and the device will fail the test. Can be set from 20-120
seconds (default is 20 seconds).
REV – 1.0
[34-2900-0513-2 ]
13 |