Inserter Disassembly - Zimmer Biomet Arcos Modular Femoral Revision System Surgical Technique

Table of Contents

Advertisement

Figure 52

Inserter Disassembly

To loosen the strike plate, use the anti-rotation handle
to hold the neck of the implant. Turn the torque
wrench counter-clockwise and depress the button
at the top of the proximal/distal inserter handle and
unthread the strike plate (Figure 52).
Torque Wrench
Proximal/Distal
Inserter Handle
Anti-rotation Handle
Arcos Broach/Calcar Proximal Body Surgical Technique
25 |
Figure 53
Unthread the distal fastener from the distal stem
implant using the 5.0 mm hex. Disengage the
proximal/distal inserter by pulling back on the inserter
collar (Figures 53 and 54).
Figure 54