Shel lab SCO6AD Installation & Operation Manual page 51

High-heat decontamination co2 incubator
Table of Contents

Advertisement

USER MAINTENANCE
7
Compare the reference device reading with the chamber
temperature display again.
If the reference device and the chamber temperature
display readings are the same or the difference falls
within the range of your protocol, the incubator is now
calibrated for temperature.
See the next step if the readings fail to match or fall
outside of your protocol range.
8
If the two readings are not the same, and the difference still
falls outside the acceptable range of your protocol, repeat
steps 3 - 7 up to two more times.
Three attempts may be required to successfully
calibrate units that are more than ±2°C out of
calibration.
9
If the temperature readings of the incubator temperature display and the reference device
still fall outside your protocol after three calibration attempts, contact your distributor or
technical support for assistance.
-OR-
End of procedure
Reference Device
Reference Device
51 |
P a g e

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SCO6AD and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents