Program Description - Pierenkemper PierenStim Parese Manual

Device for electrostimulation
Table of Contents

Advertisement

Available languages

Available languages

Pierenkemper Operating Manual - PierenStim Parese

Program description

The PierenStim Parese provides 8 integrated programs.
Program 1
Rectangular impulse with alternating polarity
D ......Pulse width 50 ms
E .......Pause 500 ms (E = 10 x D)
Therapy session time 15 min
Usage: Treatment for non-serious paralysis
Alternative: Program 4
E
D
Program 2
Trapezoidal impulse with alternating polarity
D.......Pulse width 150 ms
A ......Rising ramp (A = 1/4 x D)
B .......Working time (B = 2/4 x D)
C .......Decreasing ramp (C = 1/4 x D)
E .......Pause 1.5 s (E = 10 x D)
Therapy session time 15 min
Usage: Treatment for more serious paralysis
Alternative: Program 5
A
B
C
E
D
Program 3
Triangular impulse with alternating polarity
D ......Pulse width 300 ms
E .......Pause 3.0 s (E = 10 x D)
Therapy session time 15 min
30