Troubleshooting; Specifications - HoMedics LEADER LDRBPS-060 Manual

Semi-automatic blood pressure monitor
Table of Contents

Advertisement

Available languages

Available languages

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn
Batteries have
on when the START/
run down.
STOP button is
pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement
Cuff has been placed
error symbol shown
incorrectly.
on display or the
blood pressure value
Did you talk or move
is displayed exces-
during measurement?
sively low (or high).
Shaking of the arm
with the cuff on.
E1 error symbol
Air circuit abnormality.
shown on display
Cuff tube may not be
plugged into monitor
correctly.
E2 error symbol
Inflation pressure
shown on display
exceeding 300 mmHg.
E3 error symbol
Error determining
shown on display
measurement data.
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
29
CORRECTION
Replace them with
four new AA alkaline
batteries.
Re-insert the batteries
in the correct posi-
tions.
Wrap the cuff prop-
erly so that it is posi-
tioned correctly.
Measure again.
Keep arm steady dur-
ing measurement.
Check cuff connection.
Measure again.
Turn the unit off, then
measure again.
Measure again.

SPECIFICATIONS

Power Source:
Four 1.5V DC (AA) alkaline batteries
Measurement
Oscillometric
Method:
Pressure: 40~280 millimeters mercury (mmHg)
Measurement Range:
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
Accuracy:
Pulse: ±5% of reading
Pressure Sensor:
Semi-conductor
Inflation:
Manual Inflation
Deflation:
Automatic Pressure release valve
Memory Capacity:
60 memories
1 minute after last button operation
Auto-shut-off:
Operation
Temperature: 50°F~104°F (10°C ~40°C)
Environment:
Humidity: 15-90% R.H.
Temperature: -4°F~158°F (-20°C~70°C)
Storage Environment:
Humidity: Less than 90% R.H.
Weight:
0.43 lbs/193 grams (without batteries)
Standard size adult arm cuff:
Arm Circumference:
9"–13" (23 – 33 cm)
5"(L) x 4.1"(W) x 2.7"(H)
Dimensions:
126mm(L) x 105mm(W) x 68mm(H)
Internally Powered equipment
Type BF Equipment
IPx0-No special protection against the ingress
of water
Not suitable for use in presence of flammable
anesthetic mixture with air, oxygen or nitrous
oxide.
Continuous operation with short-time loading.
Note: These specifications are subject to change without notice.