Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MAICO MA 27/
specified below. The customer or the user of the MA 27/
used in such an environment.
Immunity Test
IEC 60601 Test Level
Conducted
3 Vrms 150 kHz to 80
RF
MHz
IEC 61000-4-
6
3 V/m 80 MHz to 2.5
GHz
Radiated RF
IEC 61000-4-
3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
8104818 Rev.2.0
Operation Manual MA 27 / MA 27e
is intended for use in the electromagnetic environment
MA 27e
MA 27e
Comp
Electromagnetic Environment-
liance
Guidance
Portable and mobile RF
3
communications equipment should
Vrms
be used no closer to any part of the
MA 27/
MA 27e
than the recommended separation
3 V/m
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.17√��
d = 1.17√�� 80 MHz to 800 MHz
d = 1.17√�� 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field Strengthens from fixed RF
transmitters, as determined by an
electromagnetic site survey
should be less than the compliance
level in each frequency range
Interference may occur in the
vicinity of equipment marked:
41
should assure that it is
, including cables
(a)
,
(b)
04/16
.