Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The MAICO MA 27/
specified below. The customer or the user of the MA 27/
used in such an environment.
Immunity Test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical Fast
Transient/Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage Dips, Short
Interruptions and Voltage
Variations on Power
Supply Lines
IEC 61000-4-11
8104818 Rev.2.0
Operation Manual MA 27 / MA 27e
is intended for use in the electromagnetic environment
MA 27e
IEC 60601 Test
Level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
±1 kV differential
mode
±2 kV common
mode
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
5% UT
(>95% dip in UT)
for 5 sec
MA 27e
Complianc
e
±6 kV
contact
±8 kV air
±2 kV for
power
supply
lines
±1 kV for
input/outp
ut lines
±1 kV
differential
mode
±2 kV
common
mode
<5% UT
(>95% dip
in UT) for
0.5 cycle
40% UT
(60% dip
in UT) for
5 cycles
70% UT
(30% dip
in UT) for
25 cycles
5% UT
39
should assure that it is
Electromagnetic
Environment-
Guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material the
relative humidity
should be greater
than 30%.
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment.
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment.
Mains power quality
should be that of a
typical commercial,
hospital, or residential
environment. If the
user of the MAICO
MA 27/
MA 27e
requires continued
operation during
power mains
interruptions, it is
recommended that
the MA 27/
MA 27e
04/16
be

Hide quick links: