Chapter 5: Site Preparation – Environmental Requirements: Humidity Requirements
Humidity
Requirements
© Agilent Technologies 1989–2002
The 3070 / 79000 family board test system is designed
to operate in the range from 5 percent to 80 percent
relative humidity (non-condensing).
If the system is subjected to condensation, as if moved
from a cold loading dock into a warm (and damper)
environment, allow at least 24 hours for the system to
recover before powering up.
Agilent 3070 / 79000 Site Preparation
5-5