1)

Getting to know the STIMPOD NMS460

1.1)

Device Description

The STIMPOD NMS460 is a Low Frequency Neuromodulation Transcutaneous Electrical Nerve Stimulation device, used for symptomatic relief and management of
chronic intractable pain and/or as and adjunctive treatment in the management of post-surgical pain, post traumatic acute pain, as well as and adjunct for pain
control due to rehabilitation.
The STIMPOD NMS460 offers two types of waveforms for the treatment of pain. The first is a Monophasic Square Wave, which is typical of normal TENS machines.
The second waveform is a Hybrid RF waveform which consists of a Monophasic Square Wave with a superimposed Radio Frequency waveform. This waveform is
proprietary and is unique to the STIMPOD NMS460. For more information on the benefits of using this modality please visit www.xavant.com.
1.2)

Accessories

WARNING:
Use of cables or other accessories other than those supplied with the STIMPOD NMS460 may result in serious injury.
NOTE:
Standard ECG Electrodes can be used for the anode. Such ECG electrodes need to conform to ANSI/AAMI EC12 on disposable ECG electrodes, ISO 10993 on
biocompatibility and the European Medical Device Directive. ECG electrodes are not included in this package.
Carry Case
WARNING:
If battery acid has leaked into the device essential circuitry may have been compromised. In the event of leakage the device must be returned to its
manufacturer for safety checks and possible repairs.
Probe Cable:
The red (anode) connector is designed to clip on to a standard ECG electrode.
The ergonomically designed Probe is used to stimulate the affected area.
Batteries:
This unit uses 4 x AAA penlight batteries.
Alkaline batteries are recommended.
When the battery is depleted the unit will prompt the user to replace the battery and switch off.
Switch device off before replacing battery.
Remove battery if the device is not to be used for an extended period of time, to prevent
leakage.