portable and mobile RF communications equipment and the STIMPOD NMS460
The STIMPOD NMS460 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the STIMPOD NMS460 can
help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the STIMPOD NMS460 as
recommended below, according to the maximum output power of the communications equipment.
Rated maximum output power
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency
of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Guidance and manufacturers declaration – electromagnetic immunity – for equipment and systems that are non-
The STIMPOD NMS460 is intended for use in the electromagnetic environment specified below. The customer or the user of the STIMPOD NMS460 should ensure that it is used in such
an environment
Immunity Test
IEC 60601 test level
Radiated immunity
80MHz – 1GHz @ 3V/m &
80MHz - 2.5GHz
10V/m
1GHz – 2.5GHz @ 10V/m
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
Not applicable
-
-
-
-
-
life supporting
Compliance level
80MHz – 1GHz @ 3V/m &
10V/m
1GHz – 2.5GHz @ 10V/m
vi
m
80 MHz to 800 MHz
d = 1,2
0,12
0,38
1,2
3,8
12
Electromagnetic environment - guidance
Portable and mobile RF communications equipment can affect MEDICAL
ELECTRICAL EQUIPMENT and should be used no closer to any part of the
equipment, including cables, than the recommended separation distance.
800 MHz to 2,5 GHz
d = 2,3
0,23
0,73
2,3
7,3
23