Labkotec idOil-OIL Installation And Operating Instructions Manual page 11

Oil sensor
Table of Contents

Advertisement

Installation and Operating Instructions | idOil-OIL
1
2
3
4
5
6
7
8
Sensor connector OilSET-1000
9
=
SHD
10
=
+
11
=
-
Shield is not connected to the control unit. Extra wires and shield must be cut and
isolated carefully.
Junction box LJB2-78-83 for a single sensor. Extra wires and possible shield are
connected to the earthing connector of the junction box.
Cable connector LCJ1-1 for a single sensor. Extra wires and shield must be cut
and isolated carefully.
Sensor connector SET-2000 Hi Level/Oil
12
=
SHD
13
=
+
14
=
-
Shield is not connected to the control unit. Extra wires and shield must be cut and
isolated carefully.
Junction box LJB3-78-83 for two sensors. Extra wires and possible shield are
connected to the earthing connector of the junction box.
Cable connector LCJ1-2 for two sensors. Extra wires and shield must be cut and
isolated carefully.
Extension cable, e.g. protected, twisted-pair 2 x 0.5 mm
68 Ohm.
idOil-OIL sensor
11/18
2
, maximum resistance
DOC001799-EN-1