Emergency Stop Button - lopital Marina Instructions For Use Manual

Shower stretcher
Table of Contents

Advertisement

Connection for
manual control

Emergency Stop Button (61002350)

• An emergency stop button is positioned in the centre of the Marina's base frame on the
transverse beam.
• In the event of an emergency, press the emergency button; the raising or lowering
movement will then stop.
• Once the emergency button has been pressed, all electric controls will stop.
• To recommence operation, turn the emergency button on the control a quarter of a turn
to the right, to unlock this.
Emergency Lowering Function (61002350)
If the manual control is no longer functioning and the trolley is in a high position,
the trolley can be lowered using the emergency lowering button.
1. Disconnect the plug from the manual control
2. Press the emergency lowering button
The trolley will now lower to the lowest position (make sure that there are no people or
objects underneath the trolley).
Remove the person from the trolley and do not use the trolley again until this has been
repaired.
Contact Lopital Nederland's technical team or service department.
Emergency lowering button
Emergency stop button