4. Safety functions

4.1. Lock button

The Reflex is fitted with a Lock button (1) (See figure 23) .
Operate the Lock button if the Reflex is not to be used for an
extended period . This will prevent any unauthorised use of
the chair .
When you press the Lock button, the red LED (2) next to the
Lock button will come on to indicate that the chair is not
working .
In order to use the chair again, you must reset it . You do this
by keeping both upwards buttons pressed in for 4 seconds .

4.2. Sleep setting

After around an hour, the chair will fall into a 'sleep' setting and the battery indication
LEDs on the handcontrol will go out . In order to use the chair again, it must be activated
(See paragraph 2.2 Activating) .

4.3. Emergency lowering

If the controls stop working and the chair is in a raised
position, it can be lowered using the emergency lowering
handle .
Check first to make sure the chair is not in sleep setting!
• Place the emergency lowering handle at the appropriate
place (1) (See figure 24) .
• Turn the handle clockwise (2) (See figure 24) .
The chair will now return to the lowest position (make sure no persons or objects are
under the chair) .
Take the client from the chair and make sure the device is not used until it has been
repaired . Place the chair in a safe place and indicate that it is out of order .
Reflex Shower-toilet chair | Instructions for use | 41
1
2
Figure 23 Lock button .
1
2
Figure 24
Emergency lowering handle .

Advertisement