T_COM_HAL_COM_ERR
FIRMWARE_SYSTEM_IN_WARMUP
FIRMWARE_MAX_SHUTTERS_OPEN
FIRMWARE_NECSEL_CONTROLLER_TEMP_SERVO
FIRMWARE_NECSEL_CONTROLLER_LASER_UNDER_TEMP
FIRMWARE_NECSEL_CONTROLLER_LASER_OVER_TEMP
FIRMWARE_NECSEL_CONTROLLER_CURRENT_AT_RAIL
PC_APP_WRITE_COMMAND_ERR
15000-MNL-01 Rev A
-20035
HAL/low level
communications failure.
Attempted to go to RUN
-20100
mode when system still in
WARMUP
-20103
Attempted to open a
shutter when maximum
number already open
-20259
Necsel Controller
temperature servo error
-20262
Necsel Controller laser
under temp error
-20263
Necsel Controller laser
over temp error
-20265
Necsel Controller current
at rail error
-40006
LDI UI - communication
error with LDI
the error persists, please contact a
technical representative.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Wait for the system to exit WARM-UP
mode and transition to RUN mode.
Only 4 shutters can be open at one time.
If this fault persists while less than four
shutters are open, unplug the unit and
wait 10 seconds before reapplying power
to the system.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Unplug the unit and wait 10 seconds
before reapplying power to the system. If
the error persists, please contact a
technical representative.
Close all shutter and let the system cool
down. If the error persists, verify that the
unit is operating within the specified
ambient temperature range and nothing
is blocking the air inlet or exhaust.
Please contact a technical representative
if this error persists.
Make sure the LDI is powered on. Check
USB cable connection to the host PC and
the LDI. Disconnect the LDI from the PC,
31